94
────────────────────────────────────────────────────
7.4 Command
────────────────────────────────────────────────────
Sets the upper level value.
Syntax
:CONF:CUPP <data>
data
0.1 - 9.9 ( Numerical data in NR2 format, two digits )
10 - 120 ( Numerical data in NR1 format, two or three digits )
Upper level value ( Refer to section 7.5. )
Response
OK
Function
In the READY or SETTING state, set the upper level value.
Example
Sets the upper level value to 5.0 mA.
PC
>
3158
>
:CONF:CUPP 5.0
OK Setting completion.
Error
In states other than the READY and SETTING states, or when a value below the
comparative-voltage value is set, an execution error occurs.
Note that an execution error occurs when setting the test upper limit to a value less
than the test lower limit when the latter is on.
Queries the upper level value enablement.
Syntax
:CONF:CUPP?
Response
<data>
data
Upper level value ( Refer to section 7.5. )
Function
Queries the upper level value enablement.
Example
PC
>
3158
>
:CONF:CUPP?
5.0 Upper level value is 5.0 mA.
:CONF:CUPP
:CONF:CUPP?