7.4 Command
────────────────────────────────────────────────────
Sets the lower level value.
Syntax
:CONF:CLOW <data>
data
0.1 - 9.9 ( Numerical data in NR2 format, two digits )
10 - 120 ( Numerical data in NR1 format, two or three digits )
Lower level value ( Refer to section 7.5. )
Response
OK
Function
In the READY or SETTING state, set the lower level value.
Example
Sets the lower level value to 0.1 mA.
PC
>
3158
>
:CONF:CLOW 0.1
OK Setting completion.
Error
In states other than the READY and SETTING states, or when a value below the
comparative-voltage value is set, an execution error occurs.
Note that an execution error occurs when setting the test lower limit to a value more
than the test upper limit when the former is on.
Queries the lower level value enablement.
Syntax
:CONF:CLOW?
Response
<data>
data
Lower level value ( Refer to section 7.5. )
Function
Queries the lower level value enablement.
Example
PC
>
3158
>
:CONF:CLOW?
0.1 The lower level value is 0.1 mA.
:CONF:CLOW
:CONF:CLOW?