112
────────────────────────────────────────────────────
9.1 Basic Specifications
────────────────────────────────────────────────────
Memory function
The Comparative Voltage Value, urrent determination for upper-
and lower-level values, and test time.
Storage of up to 20 test conditions allows quick switching between
test conditions for different standards.
Voltage comparator
Testing begins only when voltage is within 5% of the preset
value. (
50 V when the output-voltage 1 kV) The voltage
comparator prevents inadvertent testing with incorrect voltage
settings. (Can be changed to check voltage immediately prior to the
end of a test, depending on optional settings.)
Buzzer settings
Buzzer volume settings for test screening and errors (buzzer volume
settings are to be performed for PASS and FAIL separately)
Hold function
Enable this function to hold the current state when testing is
interrupted by the
STOP
key.
Pass hold function
Enable this function to hold the Pass state when detected, to
facilitate confirmation.
Fail hold function
Enable this function to hold the Fail state when detected, to
facilitate confirmation.
Momentary output
This function allows current output only when the
START
key is
pressed. The
START
key on the remote control or the START
signal via external I/O has the same effect.
Double action
Enable this function to allow testing to start only when the
START
key is pressed within about 0.5 seconds after the
STOP
key.
FAIL mode
Enable this function to restrict hold release to the
STOP
key on
the main unit.
RS Command [START]
Turn this function on to enable the RS-232C START command.
Inter-lock function
Enable this function to activate the external I/O interlock terminals.
Other Functions