8.6 Message Reference
124
Select and Query the Measurement Mode Setting
Syntax
Command
:FUNCtion <RV/ RESistance/ VOLTage>
Query
:FUNCtion?
Response <RV/ RESISTANCE/ VOLTAGE>
RV
V mode
(Resistance and voltage measurement)
RESISTANCE .......
mode (Resistance measurement)
VOLTAGE .............V mode (Voltage measurement)
Example
Command
:FUNC RV
Selects the V mode.
Query
:FUNC?
Response
RV
V mode has been selected.
Set and Query the Resistance Measurement Range
Syntax
Command
:RESistance:RANGe < 0 to 3100>
Query
:RESistance:RANGe?
Response <measurement range(NR3)>
<measurement range(NR3)> = 3.0000E-3/ 30.000E-3/ 300.00E-3/
3.0000E+0/ 30.000E+0/ 300.00E+0/ 3.0000E+3
Example
Command
:RES:RANG 120E-3
Selects the most suitable resistance measurement range for measur-
ing 120 m
.
Query
Response
:RES:RANG?
300.00E-3
The current resistance measurement range is 300 m.
Note
Changing the resistance measurement range clears stored measurement data.
Set and Query the Voltage Measurement Range
Syntax
Command
:VOLTage:RANGe <-1000 to 1000>
Query
:VOLTage:RANGe?
Response <measurement range(NR3)>
<measurement range(NR3)>
=10.00000E+0/100.0000E+0/1.00000E+3
Example
Command
:VOLT:RANG 15
Selects the voltage measurement range for measuring 15 V.
Query
Response
:VOLT:RANG?
100.0000E+0
The voltage measurement range is fixed at 100 V.