8.6 Message Reference
131
8
Chapter 8 RS-232C/GP-IB Interfaces
Set and Query the Comparator Decision Tolerance Setting
(Comparator Function)
(Resistance Measurement)
Syntax
Command
:CALCulate:LIMit:RESistance:PERCent
<Tolerance (%)>
Query
:CALCulate:LIMit:RESistance:PERCent?
Response <Tolerance (%)>
<Tolerance (%)> = 0 to 99.999 (NR2)
Example
Command
:CALC:LIM:RES:PERC 0.3
Query
:CALC:LIM:RES:PERC?
Response
0.300
(Voltage Measurement)
Syntax
Command
:CALCulate:LIMit:VOLTage:PERCent
<Tolerance (%)>
Query
:CALCulate:LIMit:VOLTage:PERCent?
Response <Tolerance (%)>
<Tolerance (%)> = 0 to 99.999 (NR2)
Example
Command
:CALC:LIM:VOLT:PERC 1.538
Query
:CALC:LIM:VOLT:PERC?
Response
1.538