Command #165: Force Relay State
Force Relay State.
Request Data Bytes
Byte Format Description
0 Integer Relay selection
1 Enum Relay state
Response Data Bytes
Byte Format Description
0 Integer Relay selection
1 Enum Relay state
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1 Undefined
2 Error Invalid Selection
3 Error Parameter Too Large
4 Error Parameter Too Small
5 Error Too few data bytes received
6 Error Transmitter Specific Command Error
7 Error Write Protect Mode
8-15 Undefined
16 Error Access Restricted
17-31 Undefined
32 Error Busy
33-127 Undefined
Command #166: Force 4-20 mA - Specific Loop Current
Force 4-20mA - Specific Loop Current
Request Data Bytes
Byte Format Description
0-3 Float mA Current
Response Data Bytes
Byte Format Description
0-3 Float mA Current