Event
Number
Event List Desc.
(Short)
Event History
Description
Data Field Problem Possible Cause Recovery
Time limit too short
Change time limit in
Configuration Profile
104
TIMEOUT PNT
DIS
Timeout Run-
time Point
Disable
0.0 Operator error
Point disabled longer than time limit setting in the Configuration Profile
Press RESET to reinitialize
timer
Restore point to Monitor Mode
Time limit too short
Change time limit in
Configuration Profile
105
TIMEOUT ALM
DIS
Timeout Run-
time Alarm
Disable
0.0 Operator error
Point disabled longer than time limit setting in the Configuration Profile
Press RESET to reinitialize
timer
Restore point to Monitor Mode
Time limit too short
Change time limit in
Configuration Profile
106
POS PNT
PRESSURE
Positive Point
Pressure
Pressure in. Hg
Positive pressure between
sampling location and monitor
Point pressure above atmospheric pressure while in idle Relieve/reduce pressure
Transducer error
Calibration error Contact Honeywell Service
Defective sensor Contact Honeywell Service
Offset incorrect Contact Honeywell Service
108
FLOW ADJ ERR
HI
High Flow After
Adjustment
cc/min.
Flow cannot be adjusted below
210 cc/min.
Debris in proportional valve
Verify range of valve
Perform Auto Balance to clear
debris
Debris or defective valve Proportional valve held open Replace with 0235-1279
109 CC EXPIRED
Chemcassette
Passed Expiration
Date
0.0
Operator error Chemcassette installed past its expiration date Replace Chemcassette
Chemcassette expired Expiration date reached Replace Chemcassette
110 INVALID CC
Invalid
Chemcassette
CC Tag does not match profile
or is unreadable
Bad configuration or corrupted RFID
Check gas configuration
Replace Chemcassette
111 DACS COM FAIL
No
Communication
from PC To Az
0.0 Ethernet communications Communications to DACS interrupted longer than 20 seconds
Check Ethernet cable at rear of
analyzer
Check Ethernet hub
connection and operation
Check Ethernet connection to
DACS
Restart DACS to clear OPC
driver problem
112
SWAPPED BAD
PUMP
Single pump
failed and
swapped
Number of bad pump -
1 or 2
Pump swap has occurred
because vacuum levels fell
below 6 in. Hg
Pump failure
Rebuild/Replace non-
operating pump
Unused slots not plugged
Install 1295-0404 connector
plug on unused slot
Large 0.018 in transport flow orifices installed
Replace all orifices with 0.015
in orifices (p/n 1290K0009 per