Event
Number
Event List Desc.
(Short)
Event History
Description
Data Field Problem Possible Cause Recovery
Analyzer)
Check valve on inactive pump leaking Replace check valve
Contact Honeywell Service
113 OPTICS MAINT
Optics Drive
High-Cleaning
Req
Drive
LED drive required to set zero
too high
Clean optics
Contact Honeywell Service
Tape leader installed
improperly
Leader not centered (Pts 1 or 5 error)
Reload Chemcassette and
recalibrate using leader
Autozero performed while tape on light or dark gray portion of leader
Reload Chemcassette and
recalibrate using leader
Dirt in optics Clean optics block
114 POINT VAC ERR
Excessive Point
Vacuum
in. Hg
Sample vacuum is greater than
5 in. Hg vacuum for longer
than 5 seconds
End of line filter clogged
Replace clogged end of line
filter
Sample line kinked
Isolate by disconnecting
possible crimped sample line
Maximum line length exceeded on a shared sample line
Isolate shared sample line
Review limits number of
analyzers, tubing ID vs. Length
(See "Nominal Transport Times"
on page168 for more
information.)
Crimped tube in cable carrier
Identify crimps in tubing
harness by checking pressure
with analyzer closed vs. open
115
AZ SW
DIAGNOSTIC
Az Software
Diagnostic
0.0
Analyzer failed to enter monitor
after pump swap
Control related fault is issued because of an improper control response
from another analyzer
Verify all analyzers enter
monitor mode successfully
116
PUMP OVER-
TEMP
Pump Over
Temperature
0.0
Pump module Temperature
exceeded
Clogged Filter Replace air filter
Fan Failure Check fans in pump module
Line voltage less than 208 VAC Verify main line voltage
117
PUMP EXHAUST
HI
High Exhaust
Pressure
0.0
Exhaust Manifold pressure
greater than 9 in. Hg (4.4 PSI)
Kinked exhaust
Check exhaust tubing for kinks
or restrictions
Exhaust tubing length exceeds 50 ft. (15 m.)
Reroute to reduce length or
change to a larger diameter
sample line
118 POWER SUPPLY
Power Supply
Fail/missing
0.0 Redundant supply not available Failure
Replace with supply 0060-
0020
119
BAD OPTICS
TEMP
Optics
Temperature Out
Of Range
Sum of error codes 1.0
to 64.0
1.0: Front block reads <0˚C Cold Environment Relocate Vertex M
2.0: Front block reads >60˚C Electronic problem Replace front optics block
4.0: Front block reads 45-60˚C
Cooling air failure Change air filter, replace fan
Hot environment Relocate Vertex M
8.0: Rear block reads <0˚ Cold environment Relocate Vertex M

Table of Contents