EasyManua.ls Logo

Huawei NetCol5000-A050 - Electric Control Box Layout for the Indoor Unit

Huawei NetCol5000-A050
360 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
the value measured by the sensor is 22°C, the sensor has an error of 2°C, and the
calibration value should be set to –2°C.
Procedure
Step 1 On the home screen, choose Maint > Sensor Calibration.
Step 2 Fill the calibration value based on the measured error.
----End
12.15 How Do I Change the Password?
Prerequisites
To modify a password, log in as the admin user or operator user.
You can only change the password of the current user.
Procedure
Step 1 On the home screen, choose Settings > User Settings to enter the User Settings
screen, as shown in Figure 12-19.
Figure 12-19 User Settings
Step 2 Tap Password to enter the password changing screen, as shown in Figure 12-20.
NO TE
It is recommended that the password should contain at least two types of characters.
NetCol5000-A050 In-row Air Cooled Smart Cooling
Product
User Manual 12 FAQ
Issue 08 (2021-09-17) Copyright © Huawei Technologies Co., Ltd. 305

Table of Contents

Other manuals for Huawei NetCol5000-A050

Related product manuals