5 | Installation INFICON
20 / 104 Ecotec E3000-Operating-instructions-kina22en1-28-(2211)
NOTICE
Property damage due to transport protection that has not been removed
The transport restraint blocks the mechanical system in the device.
Remove the transport restraint before start-up.
The transport protection is on the bottom of the Ecotec E3000 and consists of a yellow
star screw.
Fig.5: Unscrew the yellow transport restraint before the start-up
5.2 Connecting the sniffer line
NOTICE
Material damage due to a missing sniffer line
The device must not be operated without a connected sniffer line in order to avoid
overpressure in pump and measurement system.
Connect the sniffer line before you start up the device.
Do not replace the sniffer line while the device is in operation.
Align the red marking on the sniffer line plug with the red marking on the socket. Push
the sniffer line plug into the socket on the device until it locks into place.
Remove the plug by pulling the grooved ring on the plug. The ring opens the lock and
you can now pull the plug out.
5.2.1 Replacing the capillary filter of the sniffer probe
The metal capillary filter is the standard filter. With the plastic capillary filter there is
less danger of scratching the surfaces to be sniffed. The water conservation tip is
used if there is a danger of suctioning in liquids.

Table of Contents

Other manuals for Inficon Ecotec E3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon Ecotec E3000 and is the answer not in the manual?

Inficon Ecotec E3000 Specifications

General IconGeneral
BrandInficon
ModelEcotec E3000
CategorySecurity Sensors
LanguageEnglish

Related product manuals