INFICON Maintenance | 8
Ecotec E3000-Operating-instructions-kina22en1-28-(2211) 71 / 104
8 Maintenance
8.1 Calling up and managing maintenance
information
In the main menu you can call up information regarding the last operational states of
the device and its maintenance by selecting “History & maintenance.” Here you will
find especially the list of the last error messages. You will also determine which
maintenance work has to be confirmed and confirm maintenance.
Fig.39: Selection options in “History & maintenance”
Error list display The error list includes the errors and warnings that occurred during the operation of
the Ecotec E3000. Date and time are shown followed by an identification for the error
or the warning (E… with errors and W… with warnings), with a brief description of the
error or the warning.
Fig.40: List with errors and warnings
To have the full warning or error message displayed, select the appropriate line and
press the button with the magnifying glass. Detailed information for possible causes is
also specified for many messages.
Calibration list This list shows all calibrations that were carried out during the operation of the Ecotec
E3000. Information provided includes
• Date and time
• Type of calibration (internal or external)
• Number of gases (only with external calibration)