Ecotec E3000-Operating-instructions-kina22en1-28-(2211) 57 / 104
Fig.38: I•Guide information page
If there is an error message or a warning then this will be displayed instead of the
I•Guide program information.
6.5 Idle state (sleep)
The idle state is a meaningful alternative to switching the device off only if you allow
the device to run up automatically with the wake-up function, see "Miscellaneous
[}32]."
If you press the SLEEP button in the main menu of the device, the device goes into
idle state. The mass spectrometer is switched off and the pumps are stopped.
“Sleep” is replaced with “START” in the main menu and the Ecotec E3000 is run up
again with “START” or the wake-up function.
The electrical components cannot maintain operating temperature in idle mode. After
a restart, precise measurements are therefore possible only after the warm-up time,
see "Calibrate [}43]."
6.6 SERVICE
The service menu is password-protected. Settings in the service menu can be made
only after a special INFICON service training.
6.7 Retrieve information about the device
You can call up all information about the device by selecting “Info” from the main
menu. The information is spread out over nine pages. The button on the bottom right
switches you to the next page and the button on the bottom left takes you back one
page. The page number is displayed in the upper right hand corner.
Menu item Format Description
Page 1: General information
Foreline pressure mbar
Flow sccm Flow through the sniffer
line