INFICON Operation | 6
Ecotec E3000-Operating-instructions-kina22en1-28-(2211) 39 / 104
Control location • Local
• RS-232
• Local and RS-232
Local:
Only measurements values can be read out via the RS-232 interface. They are not
available to control the device.
RS-232:
The Ecotec E3000 is controlled almost exclusively via the interface. The display
serves only for a visual check. Some settings can be changed on the device. Please
use the protection via an access PIN if all functions on the device are to be
inaccessible, see "Vacuum & access control [}36]."
Local and RS-232:
The Ecotec E3000 can be controlled via the interface and via inputs on the device.
Recorder output >
Recorder scaling
• linear
• logarithmic
The output is on channel 1 (pin 1 of the I/O connection).
Recorder output >
Recorder gas
• gas 1 to 4
• auto
Set PLC > Define PLC
inputs (outputs)
• You will assign different commands to the pins of the I/O connection here.
RS-232 settings • Baud rate 1200/2400/4800/9600/19200/38400
• Blank flange LF/CR/CR + LF
RS-232 protocol • LD
• ASCII
• Diagnosis
• Printer Auto
• Printer Manual
ECO-Check If you do not use a ECO-Check calibrated leak, you should choose “Deactivate” here.
Otherwise, warning 71 “No communication with ECO-Check” will be issued at every
start of the Ecotec E3000.
If you use an ECO-Check calibrated leak, you can configure a warning here with
respect to usage duration: 14/30/60/90days.
6.4 Settings for the measurements
With delivery, the data are programmed for the following gases and the measurement
results are shown below each other in the display:
• R134a
• R32