5 | Installation INFICON
24 / 104 Ecotec E3000-Operating-instructions-kina22en1-28-(2211)
5.4 Connecting the external display unit to the
Ecotec E3000RC
Connect the external display unit and the Ecotec E3000RC with the appropriate
connecting cable. Secure the plug on the socket by tightening the screws.
Fig.9: Ecotec E3000RC with external display unit as a table device (left), for rack mounting
(right)
5.5 Connecting IC1000/BM1000
An IC1000 and various bus modules are available as accessories for the Ecotec
E3000, see "Accessories [}85]."
Refer to the interface description for details on the connection.
Required settings:
Control location: RS232 / local and RS232
Protocol: LD
Baud rate: 38400
5.6 Connecting to the power supply system
The supply voltage of the Ecotec E3000 is specified on the labeling of the mains plug
(rating plate). The Ecotec E3000 cannot be switched for other supply voltages.
WARNING
Danger from electric shock
Improperly grounded or fused products may be dangerous to life in case of a fault.
The use of the device is not permitted without a connected protective conductor.
Only use the 3-wire power cord provided.
Make sure that the power supply plug is always accessible.
Connect the device to the electric power supply using the supplied power cable.

Table of Contents