6 | Operation INFICON
50 / 104 Ecotec E3000-Operating-instructions-kina22en1-28-(2211)
6.4.6 Measure
WARNING
Risk of electric shock
Electrical voltages can be transmitted via the sniffer tip and can cause property
damage or personal injury.
► Do not touch live parts with the sniffer probe.
► Disconnect electrically operated test objects from the mains before starting the leak
test and secure them against a restart without authorization.
WARNING
Risk of eye damage
LEDs generate a bundled light that can damage your eyes.
► Do not look into the LEDs from a short distance or for longer periods of time.
CAUTION
Risk of electric shock
Sucked up liquids can trigger short circuits and cause property damage or personal
injury.
► Do not suck up liquids into the device.
► Use the water protection tip in humid environments.
NOTICE
Material damage due to a missing sniffer line
The device must not be operated without a connected sniffer line in order to avoid
overpressure in pump and measurement system.
► Connect the sniffer line before you start up the device.
► Do not replace the sniffer line while the device is in operation.
Requirements A measurement requires:
A sniffer line is connected to the basic unit.
The device has started up and is warmed up, see "Switching on [}26]."
The device is calibrated, see "Calibrate [}43]."
You have configured the device settings required for measurement, see "Basic
settings [}32]."
You have configured the measurement settings required for your measurement, see
"Settings for the measurements [}39]."