6 | Operation INFICON
26 / 104 Ecotec E3000-Operating-instructions-kina22en1-28-(2211)
6 Operation
6.1 Switching on
Connect a sniffer line and switch on the device using the mains plug.
The Ecotec E3000 starts a self-test lasting several minutes. The display shows the
heading “RUN-UP” and the individual steps of the self-test.
Fig.10: The device starts up
After running up, the Ecotec E3000 already measures the gas concentration in the
environment. There is no separate Start function. However, you still have to calibrate
the device and make different settings for your intended measurement.
If the ECO-Check calibrated leak is not in the Ecotec E3000, an acoustic warning
signal and the warning 71 are output during the first start-up.
To stop the alarm quickly, press the button on the bottom right (labeled “OK“). If you
are working without the ECO-Check, you should deactivate the alarm permanently,
see "Connecting the ECO-Check calibrated leak [}23]."