EasyManuals Logo

Inficon Ecotec E3000 User Manual

Inficon Ecotec E3000
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
INFICON Operation | 6
Ecotec E3000-Operating-instructions-kina22en1-28-(2211) 33 / 104
Japanese (Katakana)
Chinese (Mandarin, simplified Chinese)
To set the language to English temporarily, press buttons two and six during the run-
up of the Ecotec. Call up the language setting after the run-up and set the required
language.
Date & time First page: Date in the DD.MM.YYYY format
Second page (press button on the bottom right ) time with the format SS:MM.
Sniffer light Activate/deactivate light
Adjust brightness from 1 (min) to 6 (max)
Pressure unit atm
Torr
Pa
mbar
Leak rate filter Auto
Fixed
I-Filter
The I•Filter is an intelligent filter algorithm that delivers the best results regarding
interference suppression and stability of the leak rate signal. It was especially
developed for the use in the Ecotec E3000.
Only in cases in which the older Ecotec II model was replaced with an Ecotec E3000
and the Ecotec E3000 is used in a fixed test device can it be necessary to select the
older filter settings “Auto” or “Fixed.”
Alarm delay With very unstable surface conditions, it can be expedient to emit an acoustic alarm
only if the trigger value is exceeded for a specified period of time. If the function is
activated, the number indicating the volume of the device speaker flashes in the status
bar.
Setting range: 0 to 9.9seconds in steps of tenths of a second Although the alarm
delay is activated, the device may not warn if it finds a leak. Proceed as follows to
ensure that every test is successful.
1 Please note the measurement display in the device and the handle.
2 Deactivate the function when surface conditions are stable again.
Wake up If the Ecotec E3000 is in an idle state (sleep), a time can be set for it to start
automatically. This way you can set the Ecotec up to run through its warm-up phase
already before the shift starts.
You can set a separate wake-up time exact to the minute for each day of the week.
To deactivate the wake up again, set the wake-up time to 00:00.

Table of Contents

Other manuals for Inficon Ecotec E3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon Ecotec E3000 and is the answer not in the manual?

Inficon Ecotec E3000 Specifications

General IconGeneral
BrandInficon
ModelEcotec E3000
CategorySecurity Sensors
LanguageEnglish

Related product manuals