Docu-No.: 0590020601-EN Rev. 1.6.0 Page 33
Programming Reference 1
Scale menu 1.6
1.6.3 Manual addressing of SFTs
To set the address of a newly installed SFT to a prescribed address,
follow the next procedure.
1. Select SCALE menu, SFT sub-menu.
2. View SFTs variable to verify the SFT in question is at address <0>.
3. Select <SFT SELECTED>.
4. Enter the new SFT address.
5. Select <SFT ADDRESSED>.
6. Select <On>.
7. View SFTs variable to verify the SFT is now at the desired
address.
To manually change the address of a properly installed SFT to a new
defined address, follow the next procedure.
1. Select SCALE menu, SFT sub-menu.
2. View SFTs variable to verify the SFT address to be changed is
present.
3. Select <SFT SELECTED>.
4. Enter the SFT address seen from step 2.
5. Select <SFT ADDRESSED>.
6. Select <Off>.
7. View SFTs variable to verify the SFT address is now <0>.
8. Select <SFT SELECTED>.
9. Enter new SFT address.
10.Select <SFT ADDRESSED>.
11.Select <On>.
12.View SFTs variable to verify the SFT is now at the desired
address.