8. Troubleshooting
102
When the drain pump is clogged 8-7.
The sample which can not be dissolved in the solvent, clogged the piping and pump tube.
Please remove the clogging by the following steps when the drainage volume is lowered.
1) Remove the tube.
2) Insert the dropping pipette to the tip of the pump tubing or tube , remove the
clogging.
Caution
When removing the tube, please attach the eye protection and gloves. Please
be careful because there is a possibility that the drainage jump out.
Remove