4. Method
MKC-710 49
Report parameter 4-5.
Report is set when printer is connected.
[Format]
Selection of print format:
x Short :Prints sample number, measurement date, sample size,
measurement results, titration time except measurement
condition.
x GLP :Prints all of measurement parameters and results.
x Off :No printout.
[Data List]
Selection of print out of the data list.
x Off :No printout of data list.
x On :Printout of data list.
[Graph]
Selection of graphic print together with measurement results when they are
printed out.
x Off :No printout of graph
x Form1 :The accumulated titrated water is printed in graphic form. The axis of
accumulated amount is shown in percent, and the evaporation rate
against 100% water at EP is shown on display or printed out.
x Form2 :Titration volume (electrolysis) per set time and accumulated amount are
shown in graph on display or printed out.