EasyManua.ls Logo

KEM MKC-710 - Specification

KEM MKC-710
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9. Others
112
Specification 9-3.
Specification
Contents
Type and Model
MKC-710 Karl Fischer Moisture Titrator
Measuring method
Karl Fischer coulometic Titration method
Measuring range
Water content 10Pg to 300mg
Reproducibility
within 0.3%CV (n=10)/water-standard 1mgH
2
0 *
Display resolution
0.1Pg
Control method
Constant current pulse time control
Endpoint detection
AC polarization
EP sense method
Selective drift stability or Limit measurement time
Measurement cell
2-Component or 1-Component
Titration cell
Anolyte 100mL (max 150mL)
Catholyte 5mL
Number of methods
20
Display
White LED-backlit LCD
On-screen display
Potential, titration volume/ Measurement Results/ Titration conditions /
Parameters
Japanese / English / Mandarin Chinese / Korean / Russian / Spanish
Calculation
Concentration Statistics (mean, SD, RSD) / Auto input of blank
Data memory
100 samples
GLP support
Registration of operator / Record of calibration results / Record of date for
changing reagent
External I/O
RS-232C ×2: for dot matrix printer, electronic balance,
Data Acquisition Software (SOFT-CAP)
USB ×1: for USB flash drive, thermal printer, keyboard,
barcode reader, foot switch
Ambient conditions
Temperature :5 to 35
Humidity :85%RH or below (no condensation)
Power supply
DC24V 1.9A(Main unit)
AC100 - 240V r10% 50/60 Hz (Comes with AC Adapter)
Power consumption
Approx. 20W
Dimensions
Main unit 141W)×292D)×244Hmm
Stirrer 110W)×206D)×340Hmm (not incl. Solvent Change Unit)
Weight
Approx. 3kg
CE marking
EMC : EN61326
LVD : Conforming to EN61010-1
*Per KEM standard measurement conditions and standard liquids