7. Maintenance
86
7-1-4. Replacement of the anolyte (anode reagent)
Exchange timing
x When the accumulated moisture measurement (=reagent life) has reached
1000mgH2O after the last replacement with new one
x When the surface of anolyte exceeds the upper line on cell wall when sample liquid is
discharged into the cell.
x When drift level goes up.
For reagent capacity, see Instruction of the reagent you have purchased.
Please refer to 3-3-2 Anolyte
7-1-5. Replacement of the catholyte (cathode reagent)
Exchange timing
x When the accumulated moisture measurement (=reagent life) has reached
300mgH2O after the last replacement with new one.
x When drift level goes up.
Note
Negligence of replacing cathode reagent will cause higher drift level, foreign
objects generated around the diaphragm and may lead to measurement errors.
For reagent capacity, see Instruction of the reagent you have purchased.
Please refer to 3-3-1 catholyte