3. Basic operation
22
Assemble the Titration cell 3-2.
Assemble the Titration cell
1) Put a stirrer rotor into the cell, and install the inner burette, the electrode,
desiccant tube A, the port plug and syringe inlet.
desiccant tube
electrode
syringe inlet
port plug
inner burette
2) Place the titration cell onto the titration
holder, and plug in the cable from the inner
burette and the electrode. Tighten the plug
screws firmly.
Note
Make sure to apply KF grease around glass sliding area.