1. Preparations for measurement
4
Installation and start-up 1-2.
1-2-1.Assembly of titration cell
1) Put a stirrer rotor into the cell, and install the inner burette, the electrode, desiccant
tube, the port plug and syringe inlet.
Note
Make sure to apply KF grease around glass sliding area.
Note
The seal on desiccant tube A on titration flask must be removed before
using.
Warning
When handling the inner burette, do not
hold the housing (black resin area) and sliding
area of desiccant tube in order to avoid breakage.
electrode
port plug
stirrer rotor
desiccant tube
inner burette
syringe inlet