3. Basic operation
MKC-710 35
Saving Method Conditions, Setting Up on PC
Method conditions can be saved to a USB flash drive. They can also be copied
from a USB flash drive to the titrator.
Press [MENU/HOME] key on the main screen. Select 4. Data Copy” and press [ENTER].
Select “2.Method Data” and press [ENTER].
Note
This function is not available when method contents are locked. Follow the steps
in 4-1 and unlock method first.
Select how you wish to save method conditions.
Item
Description
1. Select Data
Copies each method individually. Select a method No. of either
MKC-710 or the USB flash drive, and select the method No. of
MKC-710 or the USB flash drive to which you wish to copy. Then
execute copying.
2. All Data
Copies all data at a time. Select “MKC-710USB” or
“USBMKC-710.” Then execute copying.
[Execute]
Copies the titration method to a USB flash drive or to MKC-710 as you have desired.