1. Preparations for measurement
12
The following chloroform-containing reagents (see table below) or oil-based
samples deteriorate the drain pump.
<Do not use the standard drain pump with the following Anolyte:>
KEM
FlukaRdH& HPC
Mitsubishi
Anolyte
AO
AK
AS
Therefore, when using such reagents or samples, change the connecting of the Auto
Solvent Change Unit as follows.
<Installation (in the negative pressure)>
Connect the Drain tube to Cell to waste bottle. (Use Stopper and Long Male Union
supplied to Drain tube to Waste Bottle.
Connect the Suction tube (Bottle Cap Pump) to waste bottle.
Connect the Suction tube (Bottle Cap Pump) to magnetic stirrer back side pump IN.
Note
Please refer to a previous page for the connection of other parts.
Injection tube
Suction Tube
(Bottle Cap- Pump)
Drain tube to Cell
Waste bottle
Bottle for
Anolyte
Long Male Union
Stopper