4. Method
MKC-710 45
[Unit]
Here you select a unit used in calculation.
For calculation units and numbers, refer to“4-4. Calculation formula”.
[Weight Input]
Select how to enter weight of sample.
・Fixed : “Fixed” is concentration calculation by constant sample weight.
・Variable :“Variable” is calculation by individual sample weight at each time.
This setting is for Calc.No.2, 3, 6 only, and not available for Calc.No.1, 4, 5.
The weight is entered by [ Sample] key.
[Drift Comp.]
Selection of drift compensation:
Off :No compensation is made. Select this to know total water content
including drift.
Manual :Enter offset value. This is useful when there is much difference in
drift level between start and the end of titration.
Auto :The drift level is automatically corrected with the level at time of start.