4. Method
MKC-710 45
[Unit]
Here you select a unit used in calculation.
Note
For calculation units and numbers, refer to“4-4. Calculation formula”.
[Weight Input]
Select how to enter weight of sample.
Fixed Fixed” is concentration calculation by constant sample weight.
Variable :“Variable” is calculation by individual sample weight at each time.
This setting is for Calc.No.2, 3, 6 only, and not available for Calc.No.1, 4, 5.
The weight is entered by [ Sample] key.
[Drift Comp.]
Selection of drift compensation:
Off No compensation is made. Select this to know total water content
including drift.
Manual Enter offset value. This is useful when there is much difference in
drift level between start and the end of titration.
Auto The drift level is automatically corrected with the level at time of start.