3 Remote Interface Reference
Calibration Commands
102 E364xA User’s and Service Guide
CALibration:CURRent:LEVel
{MINimum|MIDdle|MAXimum}
This command can only be used after calibration is
unsecured and the output state is ON. It sets the power
supply to a calibration point that is entered with the
CAL:CURR command. During calibration, three points must
be entered and the low- end point (MIN) must be selected
and entered first.
CALibration:SECure:CODE <quoted new code>
Enter a new security code. To change the security code, first
unsecure the power supply using the old security code.
Then, enter the new code. The calibration code may contain
up to 11 characters over the remote interface. Refer to
“Calibration” on page 61 for more information.
CALibration:SECure:STATe {OFF|ON},<quoted
code>
Unsecure or secure the power supply with a security for
calibration.
CALibration:SECure:STATe?
Query the secured state for calibration of the power supply.
The returned parameter is 0 (OFF) or 1 (ON).
CALibration:STRing <quoted string>
Record calibration information about your power supply. For
example, you can store such information as the last
calibration date, the next calibration due date, or the power
supply’s serial number. The calibration message may contain
up to 40 characters. The power supply should be unsecured
before sending a calibration message.
CALibration:STRing?
Query the calibration message and returns a quoted string.