Remote Interface Reference 3
Calibration Commands
E364xA User’s and Service Guide 103
CALibration:VOLTage[:DATA] <numeric value>
This command can only be used after calibration is
unsecured and the output state is ON. It enters a voltage
value that you obtained by reading an external meter. You
must first select the minimum calibration level
(CAL:VOLT:LEV MIN) for the value being entered. You must
then select the middle and maximum calibration levels
(CAL:VOLT:LEV MID and CAL:VOLT:LEV MAX) for the value
being entered. Three successive values must be selected and
entered. The power supply then computes new voltage
calibration constants. These constants are then stored in
non- volatile memory.
CALibration:VOLTage:LEVel
{MINimum|MIDdle|MAXimum}
This command can only be used after calibration is
unsecured and the output state is ON. It sets the power
supply to a calibration point that is entered with the
CAL:VOLT command. During calibration, three points must
be entered and the low- end point (MIN) must be selected
and entered first.
CALibration:VOLTage:PROTection
Calibrate the overvoltage protection circuit of the power
supply. It takes about 10 seconds to execute the command.
The calibration must be unsecured and the output be opened
before calibrating the overvoltage protection circuit. The
power supply automatically performs the calibration and
stores the new overvoltage constant in non- volatile memory.
Notice that voltage calibration precedes before sending this
command.