EasyManua.ls Logo

Keysight Technologies E3646A - Interface Configuration Commands

Keysight Technologies E3646A
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Remote Interface Reference
Interface Configuration Commands
106 E364xA User’s and Service Guide
Interface Configuration Commands
Refer to also “Configuring the Remote Interface” on page 42
for more information.
SYSTem:INTerface {GPIB|RS232}
Select the remote interface. Only one interface can be
enabled at a time. The GPIB interface is selected when the
power supply is shipped from the factory.
SYSTem:LOCal
Place the power supply in the local mode during RS- 232
operation. All keys on the front panel are fully functional.
SYSTem:REMote
Place the power supply in the remote mode for RS- 232
operation. All keys on the front panel, except the Local key,
are disabled while in the remote mode.
SYSTem:RWLock
Place the power supply in the remote mode for RS- 232
operation. This command is the same as the SYST:REM
command except that all keys on the front panel are
disabled, including the Local key.
<Ctrl-C>
Clear the operation in progress over the RS- 232 interface
and discard any pending output data. This is equivalent to
the IEEE- 488 device clear action over the GPIB interface.
NOTE
It is very important that you send the SYST:REM command to place the
power supply in the remote mode. Sending or receiving data over the
RS-232 interface when not configured for remote operation can cause
unpredictable results.

Table of Contents

Related product manuals