Service and Maintenance 8
Surface Mount Repair
E364xA User’s and Service Guide 183
Surface Mount Repair
Surface mount components should only be removed using
soldering irons or disordering stations expressly designed for
surface mount components.
Use of conventional solder removal equipment will almost
always result in permanent damage to the printed circuit
board and will void your Keysight factory warranty.
Replacing the Power-Line Fuse
The power- line fuse is located within the fuse- holder
assembly on the rear panel of the power supply. Refer to
“Connecting power to the power supply” on page 9 to check
the rating of power- line fuse and replace with the correct
one for your power supply.