Remote Interface Reference 3
Introduction to the SCPI Language
E364xA User’s and Service Guide 77
Using the MIN and MAX parameters
You can substitute MINimum or MAXimum in place of a
parameter for many commands. For example, consider the
following command:
CURRent {<current>|MIN|MAX}
Instead of selecting a specific current, you can substitute
MINimum to set the current to its minimum value or
MAXimum to set the current to its maximum value.
Querying parameter settings
You can query the value of most parameters by adding a
question mark (?) to the command. For example, the
following command sets the output current to 5 A:
CURR 5
You can query the value by executing:
CURR?
You can also query the maximum or minimum value allowed
with the present function as follows:
CURR? MAX
CURR? MIN
CAUTION
If you send two query commands without reading the response from
the first, and then attempt to read the second response, you may
receive some data from the first response followed by the complete
second response. To avoid this, do not send a query command without
reading the response. When you cannot avoid this situation, send a
device clear before sending the second query command.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E3646A and is the answer not in the manual?

Keysight Technologies E3646A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE3646A
CategoryPower Supply
LanguageEnglish

Related product manuals