EasyManua.ls Logo

Keysight Technologies E3646A - Page 129

Keysight Technologies E3646A
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Interface Reference 3
Calibration Commands
E364xA User’s and Service Guide 105
13 Repeat step 5 through step 9 by substituting CURR for
VOLT for current calibration. For example,
CAL:CURR:LEV MIN
14 Repeat step 1 through step 13 for the other output
calibration.
15 Record calibration information such as next calibration
due date or contact person for future reference. The
calibration string may contain up to 40 characters.
CALibration:STRing ‘<string>’
NOTE
You should wait for the DVM reading to stabilize for accurate calibration.

Table of Contents

Related product manuals