3 Remote Interface Reference
Status Reporting Commands
120 E364xA User’s and Service Guide
STATus:QUEStionable:INSTrument:ENABle
<enable value>
Set the value of the questionable instrument enable register.
This register is a mask for enabling specific bits from the
questionable instrument event register to set the instrument
summary bit (ISUM, bit 13) of the Questionable Status
register. The ISUM bit of the Questionable Status register is
the logical OR of all the questionable instrument event
register bits that are enabled by the questionable instrument
enable register.
STATus:QUEStionable:INSTrument:ENABle?
Return the value of the questionable instrument enable
register.
STATus:QUEStionable:INSTrument:ISUMmary<n>[
:EVENt]?
Return the value of the questionable instrument Isummary
event register for a specific output of the two- output power
supply. The particular output must be specified by a
numeric value. n is 1 or 2. 1 selects output1, and 2 selects
output2. The event register is a read- only register which
holds (latches) all events. Reading the questionable
instrument Isummary event register clears it.
STATus:QUEStionable:INSTrument:ISUMmary<n>:
CONDition?
Return the CV or CC condition of the specified instrument.
If 2 is returned, the queried instrument is in the CV
operating mode. If 1 is returned, the queried instrument is
in the CC operating mode. If 0 is returned, the outputs of
the instrument are off or unregulated. If 3 is returned, the
instrument is in the hardware failure. n is 1, 2, or 3.