Measurement
157
The following three methods are available for executing the target search:
Target
search
(Search
Peak)
The marker moves to the peak with maximum response
value if the peak polarity is Positive or Both or to the
peak with minimum response value if the peak polarity
is Negative.
Search left
(Search Left)
Executes the search from the current marker position to
the smaller stimulus values and moves the marker to
first encountered target.
Search right
(Search
Right)
Executes the search from the current marker position to
the larger stimulus values and moves the marker to
first encountered target.
Procedure
1. Activate the marker you are using for the target search.
2. Press Marker Search key.
3. Click Target > Target Value.
4. Enter the target value in the entry area that appears. This causes the
target search to be executed based on the definitions of the newly
set target value and the currently set transition type.
5. Click Target Transition.