Measurement
179
two oe more limit values at the same measurement point. in
this case, the limit value to be used in the limit test is defined
as follows:
When two or more limit values of which the type is set to
maximum (MAX) exist, the smallest one is used as the
maximum.
When two or more limit values of which the type is set to
minimum (MIN) exist, the largest one is used as the
minimum.
Even if the span of the sweep range on the E4990A is set
to 0, enter the two parameters of Begin Stimulus and End Stimulus.
When two or more response values are returned as a
result of using the complex or polar format, the first response value
of the marker provides the object of the limit test.
Saving/Calling the Limit Table
You can save the limit table to a file that you can then freely bring up on
the screen later and use. You can import a file saved in CSV format
(extension: *.csv) into spreadsheet software on a PC for later use (a
numerical value is saved as strings that includes its unit).