Programming
339
Query Response
{numeric 1}, .... ,{numeric NOP×2}<newline><^END>
Examples
See :CALC:DATA:FDAT
Related Commands
:CALC:DATA:FDAT
:CALC:DATA:RDAT
:CALC:DATA:RMEM
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.DATA.FMEMORY
:CALC<Ch>:DATA:RDAT
Type
Command/Query
Syntax
:CALCulate<Ch>[:SELected]:DATA:RDATa <Value>
:CALCulate<Ch>[:SELected]:DATA:RDATa?
Description
This command sets/gets the raw data. Save/Recall valid when
MMEM:STOR:STYPE = [DSTate|CDSTate]
Variable
Parameter Value
Description
raw data array
Where n is an integer between 1 and NOP (number of
measurement points):
<numeric n×2-1>: Real part of data (complex
number) at the n-th measurement point.
<numeric n×2>: Imaginary part of data (complex
number) at the n-th measurement point. Always 0
when the data format is not the Smith chart format or
the polar format
The number of data is {NOP×2}
Data Type Variant type Array (Range)