E4990A Help
342
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.DATA.XAXIS
:CALC<Ch>:EPAR
Type
Command
Syntax
:CALCulate<Ch>:EPARameters[:EXECute]
Description
This command executes the equivalent circuit analysis in the selected
equivalent circuit model. Executes the equivalent circuit analysis in the
partial search range.
Examples
Ana.WriteString ":CALC1:EPAR", True
Related Commands
:CALC:EPAR:CIRC
Equivalent Softkey
Analysis > Equivalent Circuit > Calculate
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).EPARAMETERS.EXECUTE
:CALC<Ch>:EPAR:CIRC
Type
Command/Query
Syntax
:CALCulate<Ch>:EPARameters:CIRCuit[:TYPE] {A|B|C|D|E|F|G}
:CALCulate<Ch>:EPARameters:CIRCuit[:TYPE]?
Description
This command selects the equivalent circuit model.
Variable
Parameter Selection Option
Description Equivalent circuit model
Data Type Character string type (String)
Range A|B|C|D|E|F|G
Preset Value
A