E4990A Help
358
negative peak) within the range to the left of the resonant point
within the waveform analysis range.
• Ripple 2: the maximum value of the right ripple (the sweep
parameter difference between the peak and the right adjacent
negative peak) within the range between the resonant point and the
anti-resonant point within the waveform analysis range.
• Ripple 3: the maximum value of the left-side ripple within the range
to the right of the resonant point within the waveform analysis
range.
This searches for phase 0° points within the analysis
range from the left edge, and defines the 1st detected point as
the resonant point and the 2nd detected point as the anti-
resonant point. Even if 3 or more phase 0° points are detected,
only the first 2 points are used. If only 1 phase 0° point is
detected, 0 is read out for parameters other than Zr, fr, and
Rpl1. If no phase 0° point is detected, 0 is read out for all the
parameters.
{numeric (Zr: Resonant impedance)},{numeric (fr: Resonant frequency)},
{numeric (Za: Antiresonant impedance)}, {numeric (fa: Antiresonant