EasyManua.ls Logo

Keysight E4990A - Page 380

Keysight E4990A
669 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E4990A Help
370
right adjacent negative peak).
"RESO":Specifies the resonant analysis.
"RESR":Specifies the resonant analysis with ripples.
"CERR":Specifies the ceramic oscillator analysis.
Preset
Value
PTPeak
Query Response
{PTP|STDEV|MEAN|MAX|MIN|PEAK|APE|ATAR|SDEV|RPP|RPLH|RPLL|RPLR
|RESO|RESR|CERR}<newline><^END>
Examples
Dim Var as String
Var= "PTP"
Ana.WriteString ":CALC1:FUNC:TYPE " & Var, True
Ana.WriteString ":CALC1:FUNC:TYPE?", True
Var=Ana.ReadString
Related Commands
:CALC:FUNC:DATA
:CALC:FUNC:EXEC
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.FUNCTION.TYPE
:CALC<Ch>:LIM
Type
Command/Query
Syntax
:CALCulate<Ch>[:SELected]:LIMit[:STATe] {ON|OFF|1|0}
:CALCulate<Ch>[:SELected]:LIMit[:STATe]?
Description
This command turns the limit test On/Off.
Variable
Parameter Selection Option
Description Limit test On/Off status

Table of Contents

Other manuals for Keysight E4990A