E4990A Help  
372 
  Data(n*5-4) :The type of the n-th line. 
Specify an integer 0 to 2 as follows. 
0: OFF 
1: Upper limit line 
2: Lower limit line 
  Data(n*5-3) :The value on the horizontal axis 
(frequency/osc level/DC bias) of the start point of 
the n-th line. 
  Data(n*5-2) :The value on the horizontal axis 
(frequency/power/DC bias) of the end point of the 
n-th line. 
  Data(n*5-1) :The value on the vertical axis of the 
start point of the n-th line. 
  Data(n*5) :The value on the vertical axis of the end 
point of the n-th line. 
The index of the array starts from 0. 
Data Type  Variant type Array (Range) 
Note 
If there is no array data, an error occurs at execution and 
the object is ignored. 
For Data(n×5-4) in the array data, if you specify an 
integer other than 0, 1 or 2, an error occurs at execution. 
For Data(n×5-3), Data(n×5-2), Data(n×5-1), and 
Data(n×5) in the array data, if the specified value is out 
of the allowable setup range, the minimum value (if the 
lower limit of the range is not reached) or the maximum 
value (if the upper limit of the range is exceeded) is set. 
Query Response 
{numeric 1},...{numeric 1+(n*5)}<newline><^END> 
Examples 
See limit line. 
Related Commands 
:CALC:LIM 
Equivalent Softkey 
Analysis > Limit Test > Edit Limit Line