372
Data(0).
Data(n*5-4) :The type of the n-th line.
Specify an integer 0 to 2 as follows.
0: OFF
1: Upper limit line
2: Lower limit line
Data(n*5-3) :The value on the horizontal axis
(frequency/osc level/DC bias) of the start point of
the n-th line.
Data(n*5-2) :The value on the horizontal axis
(frequency/power/DC bias) of the end point of the
n-th line.
Data(n*5-1) :The value on the vertical axis of the
start point of the n-th line.
Data(n*5) :The value on the vertical axis of the end
point of the n-th line.
The index of the array starts from 0.
Data Type Variant type Array (Range)
Note
If there is no array data, an error occurs at execution and
the object is ignored.
For Data(n×5-4) in the array data, if you specify an
integer other than 0, 1 or 2, an error occurs at execution.
For Data(n×5-3), Data(n×5-2), Data(n×5-1), and
Data(n×5) in the array data, if the specified value is out
of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum
value (if the upper limit of the range is exceeded) is set.
Query Response
{numeric 1},...{numeric 1+(n*5)}<newline><^END>
Examples
See limit line.
Related Commands
:CALC:LIM
Equivalent Softkey
Analysis > Limit Test > Edit Limit Line