384
:CALC<Ch>:MARK<Mk>:BWID:FIX:VAL
Type
Command/Query
Syntax
:CALCulate<Ch>[:SELected]:MARKer<Mk>:BWIDth:FIXed:VALue <Value>
:CALCulate<Ch>[:SELected]:MARKer<Mk>:BWIDth:FIXed:VALue?
Description
This command sets/gets the marker bandwidth fixed value.
Variable
Parameter Value
Description Marker bandwidth fixed value
Data Type Numeric type (Real)
Range -500M ~ 500M
Preset Value
0
Unit Depends on Measurement Format
Resolution -
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Double
Var= 10
Ana.WriteString ":CALC1:MARK1:BWID:FIX:VAL " & Str(Var), True
Ana.WriteString ":CALC1:MARK1:BWID:FIX:VAL?", True
Var=Ana.ReadNumber
Related Commands
:CALC:MARK:BWID
:CALC:MARK:BWID:TYPE
Equivalent Softkey
Marker Search > BW Fixed Value
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.MARKER(Mk).BWIDTH.FIXED.VALUE