Programming
383
Ana.WriteString ":CALC1:MARK1:BWID?", True
Var=Ana.ReadNumber
Related Commands
:CALC:MARK:BWID:DATA
Equivalent Softkey
Marker Search > Bandwidth
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.MARKER.BWIDTH.STATE
:CALC<Ch>:MARK:BWID:DATA
Type
Query
Syntax
:CALCulate<Ch>[:SELected]:MARKer:BWIDth:DATA?
Description
This command reads the bandwidth search result of marker 1 to 9 (Mk)
and reference marker (Mk:10).
If the bandwidth search is impossible, an error occurs when executed and
the object is ignored.
Query Response
{numeric 1}, {numeric 2}, {numeric 3}, {numeric 4}, {numeric 5},
{numeric 6}<newline><^END>
{numeric 1} : Bandwidth
{numeric 2} : Center point frequency of the 2 cutoff frequency points
{numeric 3} : Q value
{numeric 4} : marker value (peak value)
{numeric 5} : L
{numeric 6} : R
Examples
Dim Var as Variant
Ana.WriteString ":CALC1:MARK:BWID:DATA?", True
Var=Ana.ReadNumber
Related Commands
:CALC:MARK:BWID
Equivalent Softkey
No equivalent key is available on the front panel.
(Result on the screen for Marker Search > Bandwidth)
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.MARKER.BWIDTH.DATA