E4990A Help
410
SCPI.CALCULATE(Ch).SELECTED.MARKER(Mk).X.POSITION
:CALC<Ch>:MARK:XUN
Type
Command/Query
Syntax
:CALCulate<Ch>[:SELected]:MARKer:XUNit {STIM|TIME|RFReq}
:CALCulate<Ch>[:SELected]:MARKer:XUNit?
Description
This command sets/gets the marker x-axis display method for the active
trace.
Variable
Parameter Selection Option
Description Marker x-axis display method
Data Type Character string type (String)
Range
STIM: Sweep parameter
TIME: Time
RFReq: Relaxation time
Preset Value
STIM
Query Response
{STIM|TIME|RFR}<newline><^END>
Examples
Dim Var as String
Var= "STIM"
Ana.WriteString ":CALC1:MARK:XUN " & Var, True
Ana.WriteString ":CALC1:MARK:XUN?", True
Var=Ana.ReadString
Related Commands
:CALC:MARK:ACT
Equivalent Softkey
Marker Function > X Unit
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.MARKER.XUNIT
:CALC<Ch>:MARK<Mk>:Y