E4990A Help
426
Range 0 ~ 255
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Long
Var= 0
Ana.WriteString ":CONT:HAND:E " & Str(Var), True
Ana.WriteString ":CONT:HAND:E?", True
Var=Ana.ReadNumber
Related Commands
:CONT:HAND:C
:CONT:HAND:C:MODE
:CONT:HAND:D
:CONT:HAND:D:MODE
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.CONTROL.HANDLER.E.DATA
:CONT:HAND:F
Type
Command
Syntax
:CONTrol:HANDler:F[:DATA] <Value>
Description
Outputs port information to output port F (port A + port B) of the handler
I/O. Port information is output as 16-bit binary using A0 as LSB and B7 as
MSB.
The bit 14 of the data output by this project is ignored when outputting the
INDEX signal is turned ON (specifying True with the
:CONT:HAND:IND:STAT).
The bit 15 of the data output by this project is ignored when outputting the
READY FOR TRIGGER signal is turned ON (specifying True with the
:CONT:HAND:RTR:STAT).
Variable
Parameter Value