Programming
503
Dim Var as String
Var= "A"
Ana.WriteString ":MMEM:STOR:CHAN:COEF " & Var, True
Related Commands
:MMEM:LOAD:CHAN
:DISP:WIND:ACT
Equivalent Softkey
Save/Recall > Save Channel > Cal Only A | Cal Only B | Cal Only C | Cal Only D
COM Command Syntax (Internal Control Only)
SCPI.MMEMORY.STORE.CHANNEL.COEFICIENT
:MMEM:STOR:EPAR
Type
Command
Syntax
:MMEMory:STORe:EPARameters <Value>
Description
This command saves the equivalent circuit parameters of an active channel
into a CSV file at defined location.
Specify the file name with the extension. When
you use directory names and file name, separate them
with "\" (back slash), or "/" (slash).
Variable
Parameter Value
Description Filename and destination name you want save
Data Type Character string type (String)
Range 254 chars
Examples
Dim Var as String
Var= "C:\sample.csv"
Ana.WriteString ":MMEM:STOR:EPAR " & Var, True
Equivalent Softkey
There is no equivalent key available on the front panel. However, similar
key which exports to TXT file is shown below:
Analysis > Equivalent Circuit > Export to TXT File...
COM Command Syntax (Internal Control Only)
SCPI.MMEMORY.STORE.EPARAMETERS