E4990A Help  
556 
Description 
This command returns the frequency stimulus data. 
Variable 
Parameter  Value 
Description
Indicates the array data (frequency) of NOP (number of 
measurement points). Where n is an integer between 1 
and NOP. 
Data(n-1): Frequency at the n-th measurement point 
The index of the array starts from 0. 
Data Type  Variant type Array (Range) 
Query Response 
{numeric 1}, .... ,{numeric NOP}<newline><^END> 
Examples 
Dim Var as Variant 
Ana.WriteString ":SENS1:FREQ:DATA?", True 
Var=Ana.ReadList(ASCIIType_R8, ",") 
Related Commands 
:SENS:FREQ 
Equivalent Softkey 
No equivalent key is available on the front panel. 
COM Command Syntax (Internal Control Only) 
SCPI.SENSE(Ch).FREQUENCY.DATA 
:SENS<Ch>:FREQ:SPAN 
Type 
Command/Query 
Syntax 
:SENSe<Ch>:FREQuency:SPAN <Value> 
 
:SENSe<Ch>:FREQuency:SPAN? 
Description 
This command sets/gets the frequency span in frequency sweep. 
Variable 
Parameter  Value 
Description  Frequency span