Programming
557
Data Type Numeric type (Real)
Range 0 ~ 119.99998M
Preset Value
119.99998M
Unit Hz
Resolution 1m
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Double
Var= 100000
Ana.WriteString ":SENS1:FREQ:SPAN " & Str(Var), True
Ana.WriteString ":SENS1:FREQ:SPAN?", True
Var=Ana.ReadNumber
Related Commands
:SENS:FREQ:CENT
Equivalent Softkey
Span (in frequency sweep)
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).FREQUENCY.SPAN
:SENS<Ch>:FREQ:STAR
Type
Command/Query
Syntax
:SENSe<Ch>:FREQuency:STARt <Value>
:SENSe<Ch>:FREQuency:STARt?
Description
This command sets/gets the start frequency in frequency sweep.
Variable
Parameter Value
Description Start frequency

Table of Contents

Other manuals for Keysight E4990A

Related product manuals