E4990A Help
558
Data Type Numeric type (Real)
Range 20 to Maximum Frequency
Preset Value
20
Unit Hz
Resolution 1m
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Double
Var= 1000
Ana.WriteString ":SENS1:FREQ:STAR " & Str(Var), True
Ana.WriteString ":SENS1:FREQ:STAR?", True
Var=Ana.ReadNumber
Related Commands
:SENS:FREQ:STOP
Equivalent Softkey
Start (in frequency sweep)
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).FREQUENCY.START
:SENS<Ch>:FREQ:STOP
Type
Command/Query
Syntax
:SENSe<Ch>:FREQuency:STOP <Value>
:SENSe<Ch>:FREQuency:STOP?
Description
This command sets/gets the stop frequency in frequency sweep.
Variable
Parameter Value
Description Stop frequency

Table of Contents