EasyManua.ls Logo

Keysight E4990A - Page 572

Keysight E4990A
669 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E4990A Help
562
segment. This is not necessary when<meas time
on/off>=0.
[averaging factor n]: averaging factor of the n-th
segment. This is not necessary when<list average
on/off>=0.
[segment time n]: segment time of the n-th segment.
This is not necessary when<segment time on/off>=0.
[delay time n]: delay time of the n-th segment. This is
not necessary when<delay time on/off>=0.
If the necessary amount of array data for the specified
number of segments is not available while setting the
segment sweep table, an error occurs when its executed
and the object is ignored.
Data Type Variant type Array (Range)
Preset
Value
7,0,0,0,0,0,0,0,1,20,20,2
Query Response
{numeric}, ... , {numeric}<newline><^END>
Examples
Dim Var as String
riantAna.WriteString ":SENS1:SEGM:DATA 7, 0, 1, 0, 1 ,1, 0, 0, 2, 1E6, 2E6, 10, 0, 5E-3, 1, 1, 3E6, 5E6,
20, 1 , 2E-3, 5, 10", True
Ana.WriteString ":SENS1:SEGM:DATA?", True
Var=Ana.ReadString
Start Stop Points
OSC
Mode
OSC
Level
Meas
Time
Point
Avg
1
1
MHz
2
MHz
10 Volt 5 mV 1 1
2
3
MHz
5
MHz
20 Curr 2 mA 5 10